dc.description.abstract Radiochromic film has become an important tool to verify dose distributions for intensity-modulated radiotherapy (IMRT) and quality assurance (QA) procedures. A new radiochromic film model, EBT3, has recently become available, whose composition and thickness of the sensitive layer are the same as those of previous EBT2 films. However, a matte polyester layer was added to EBT3 to prevent the formation of Newton’s rings. Furthermore, the symmetrical design of EBT3 allows the user to eliminate side-orientation dependence. This film and the flatbed scanner, Epson Perfection V750, form a dosimetry system whose intrinsic characteristics were studied in this work. In addition, uncertainties associated with these intrinsic characteristics and the total uncertainty of the dosimetry system were determined. The analysis of the response of the radiochromic film (net optical density) and the fitting of the experimental data to a potential function yielded an uncertainty of 2.6%, 4.3%, and 4.1% for the red, green, and blue channels, respectively. In this work, the dosimetry system presents an uncertainty in resolving the dose of 1.8% for doses greater than 0.8 Gy and less than 6 Gy for red channel. The films irradiated between 0 and 120 Gy show differences in the response when scanned in portrait or landscape mode; less uncertainty was found when using the portrait mode. The response of the film depended on the position on the bed of the scanner, contributing an uncertainty of 2% for the red, 3% for the green, and 4.5% for the blue when placing the film around the center of the bed of scanner. Furthermore, the uniformity and reproducibility radiochromic film and reproducibility of the response of the scanner contribute less than 1% to the overall uncertainty in dose. Finally, the total dose uncertainty was 3.2%, 4.9%, and 5.2% for red, green, and blue channels, respectively. The above uncertainty values were obtained by minimizing the contribution to the total dose uncertainty of the film orientation and film homogeneity. es
dc.language eng
dc.language.iso eng es
dc.publisher JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS es
dc.relation.ispartofseries 17(5);doi:10.1120/jacmp.v17i5.6262
dc.rights openAccess es
dc.rights.uri http://creativecommons.org/licenses/by-nc-nd/4.0
dc.subject radiochromic film es
dc.subject flatbed scanner es
dc.subject dosimetry es
dc.subject uncertain analysis es
dc.subject.classification BIOLOGÍA Y QUÍMICA
dc.title Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density es
dc.type Artículo es
dc.provenance Científica es
dc.road Dorada es
dc.organismo Medicina es
dc.ambito Internacional es
dc.cve.CenCos 10301 es
dc.cve.progEstudios 6131 es
dc.audience students es
dc.audience researchers es
dc.type.conacyt article
dc.identificator 2


Ficheros en el objeto digital

Este ítem aparece en la(s) siguiente(s) colección(ones)

Visualización del Documento

  • Título
  • Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density
  • Autor
  • CAMACHO LOPEZ, MIGUEL ANGEL
  • LEON MARROQUIN, ELSA YAZMIN
  • HERRERA GONZALEZ, JOSE ALEJANDRO
  • Villarreal Barajas, Jose Eduardo
  • GARCIA GARDUÑO, OLIVIA AMANDA
  • Fecha de publicación
  • 2016-06
  • Editor
  • JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS
  • Tipo de documento
  • Artículo
  • Palabras clave
  • radiochromic film
  • flatbed scanner
  • dosimetry
  • uncertain analysis